|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
5.5 High frequency signal or noise? Similarly, data corresponding to missed events was arbitrarily chosen from the set of data meeting the following criteria: Note: The results discussed for the three tests subjects chosen for analysis in this section were indicative of the overall test population unless specifically noted otherwise. The test subjects were chosen to demonstrate both the variability and consistence of various features which were relevant to this study. Data from each of the three subjects was analyzed and examined individually. Analysis was as follows: A (windowed) power spectrum (PSD) was computed for each of the 45 data segments. The PSD's corresponding to the 25 hit events were averaged and the PSD's for the 20 missed events were also averaged, resulting in two spectral plots. These averaged spectrum provided a convenient mechanism to examine the changes in spectral components related to "typical" hit and miss events for each subject. Note: Because, at various scales in data segments, the 60Hz and subsequent harmonic components (especially the odd harmonics) tend to dominate the local scale of the power spectra, steps were taken to eliminate this activity from the PSD plots. To eliminate the 60Hz components (and harmonics), Table 5.5.1 outlines the steps used. Step 1, the center frequency to be eliminated (for example 60Hz) is determined. Step 2, the location in the power spectrum of the undesirable frequencies and a small span of frequencies on either side are determined. Step 3, components are replaced in the spectrum by the value of the nearest lower frequency component.
Table 5.5.1 Procedure to Eliminate 60Hz and Harmonic The first test subject to be examined was cia01. The averaged spectra for hit and missed events is shown in Figure 5.5.1 for the classical frequency range of 0-40Hz. In Figure 5.5.1(a), we see the average spectra of the selected 25 hit events. Similarly, Figure 5.5.1(b) shows the average of the missed events (on the same scale as the hits). The energy in the delta band remained relatively constant (decreases by a factor of 1.07), theta band energy increased by a factor of 2.57 from the hits to the misses, alpha power decreased by a factor of 2.39, and beta energy remained essentially unchanged. These shifts in energy indicate (typically) relaxed wakefulness during the hits and possible stage 1 sleep for the missed events according to the R&K (1968) criteria (i.e. alpha activity being replaced by theta activity).
Table 5.5.2 Energy distribution of Fig. 5.5.1 shown as actual energy
Figure 5.5.2 shows the remainder of the power spectrum (31 to 475Hz) with the hit and miss spectra again shown on the same scale. Although there are no dominant peaks in this frequency range comparable to those seen in the 0-40Hz range (e.g. alpha peaks, etc.), we can see that the overall magnitude of the power in the high frequency band does decrease with drowsiness. When the full spectrum is viewed on a single plot, the signals in the 0-40Hz band are much stronger than all the others and consequently, the high frequency activity is virtually invisible. By viewing the various frequency bands on their own individual scales, we can see this otherwise obscured activity quite clearly. Also, by comparing the hit and miss spectra, those changes in the high frequency components which might otherwise go unnoticed are detectable. This situation is further improved because data for the hit/miss spectra are derived from data from a single test subject and useful trends are not obscured by possible person-to-person and/or day-to-day variations in activity.
Table 5.5.3 Energy distribution of Fig. 5.5.2 shown as actual energy
Figure 5.5.3 shows a closer look at the 100-475Hz band where the ordinate has to be rescaled. Because of the extreme difference in scale between the hit and miss spectra, the scale for the missed events is 10-times smaller than for the hits.
In order to investigate this localized peak, Figure 5.5.4 shows the power spectrum computed from the ambient noise data which was simultaneously collected along with the neurological data. The noise data was collected using a 10k dummy load at the EEG electrode input terminal. The dummy load will certainly have dissimilar receiver characteristics as compared to the applied EEG electrodes attached to the test subjects. But, in spite of the different characteristics, the noise channel can still be used to check for frequency content of non-neurologic origin when a questionable peak appears on an EEG power spectrum. The data from the noise channel (specifically in the 31 to 475Hz range) remains unchanged from hits to misses, providing further evidence that the shifts in energy from hits to misses are not due to external causes.. Regarding the 263.82Hz peak: because neither noise spectra revealed the peak seen in the miss spectra given above, it was suspected that the peak could be of neurological origin.
Although the averaged power spectral plots provide insight into the relationship between behavior and the EEG data, it is also useful to examine the non-averaged data segments. To further illustrate the frequency shifts associated with the hit and missed events, Figures 5.5.a-f show the energy in various frequency bands for the individual hit and missed events. The first 25 events (black) in each plot correspond to the hit events and the last 20 events (red) correspond to the misses. The events chosen here are the same events that were used to compute the averaged power plots. Figures 5.5.a-d show the energy for each of the hit and missed events in the classical frequency bands (delta, theta, alpha, and beta). The changes in magnitude corresponding to hits and misses have already been discussed using the averaged spectral plots in these energy bands. Although the aggregate behavior in each of the frequency bands is consistent, the behavior of individual events reveals a significant overlap in the energy magnitudes between the hits and misses. That is, individual hit and missed events cannot be distinguished reliably by simply looking at the magnitude of the energy in a particular band. This has also been demonstrated across subjects in this study.
But, if we examine Figure 5.5.e (and Figure 5.5.f), the change in energy from hits to misses is approximately 10.15:1 (and 18.54:1 for Fig. 5.5.f) which appears sufficient to distinguish between hit and missed events by high frequency energy measurement alone for this selection of events. This ability to distinguish hit and missed events based on changes in high frequency energy is consistent and universal among all test subjects under the criteria set forth for event data inclusion given previously in this section.
To further illustrate the high frequency activity, let's examine the another test subject (kal01) using the same procedure of creating spectral averages from 25 hit and 20 missed events. In Figure 5.5.5, the 0 to 40Hz spectral plots show the differences between hit and missed events for this new test subject in the traditional frequency range, and Table 5.5.4 gives the accompanying energy distribution values. Similar to the previous subject, an increase in theta energy accompanied the missed events by a factor of 3.83 (a value of 2.57 for the previous test subject). But, whereas the previous subject had little change in delta energy and a decrease in alpha energy by a factor of 2.39, this subject decreased in delta energy by a factor of 2.14 and had virtually no change in their alpha energy levels between the hits to misses (increased energy by a factor of 1.01). On the high frequency side, the 0 to 40Hz band during hits accounts for 94.1855% of the energy in the spectrum, but the 0 to 40Hz band during missed signals accounts for 99.5108% of the total energy. This again shows a consistent change in the high frequency energy distribution which accompanies the behavioral failures.
Table 5.5.4 Energy distribution of Fig. 5.5.5 shown as actual energy Figure 5.5.6 and accompanying Table 5.5.5 shows the remaining 31 to 475Hz frequency range. As observed in the previous subject, order of magnitude changes in the various high frequency bands decrease consistently across the high frequencies throughout the spectrum during failures.
Table 5.5.5 Energy distribution of Fig. 5.5.6 shown as actual energy Figures 5.5.g-j show the energy in the traditional frequency bands for the individual hit and missed events. The delta and theta energy bands appear to be the best indicators of the failure, but again, individual hits and missed events could not be distinguished in all instances by simply looking at the magnitude of the energy in a particular band.
Figures 5.5.k-l correspond to the high frequency energy bands. These examples demonstrate that hits and misses can be reliably distinguished simply by examining changes in the high frequency energy content.
Finally, the third test subject in the series (cvj01) is shown in Figures 5.5.7 and 5.5.8 with accompanying Tables 5.5.6 and 5.5.7, respectively. On the low frequency side, delta energy increased by a factor of 5.29, theta energy increased with missed signals by a factor of 9.46, but, this time alpha energy increased during misses by a factor of 1.70. On the high frequency side, we see that the 0 to 40Hz band accounted for 93.2165% of the total energy during hits and increases to 99.7772% during the missed signals.
Table 5.5.6 Energy distribution of Fig. 5.5.7 shown as actual energy
Table 5.5.7 Energy distribution of Fig. 5.5.8 shown as actual energy Figures 5.5m-r show the energy in some of the various frequency bands (traditional and high frequency) for the individual hit and missed events used in the averaged spectra.
Table 5.5.8 summarizes the changes in energy in specified bands for each of the three example subjects discussed. The changes are listed in the form of multipliers from hits to misses. For example, a value of -9.02 is interpreted as a decrease in energy by a factor of 9.02 from hits to misses. Similarly, a value of +9.02 would indicate an increase in energy from hits to misses.
Table 5.5.8 Summary of the Changes in Energy Distribution (multipliers)
Even though the theta energy behaved consistently for the three test subjects shown, the alpha energy during missed events was erratic with: no changes for one subject, decreased for another, and increased for yet a third. Certainly any detection system which relies on alpha energy for drowsiness detection would be confounded by the potentially inconsistent behavior of this frequency band. We have observed, in general, that alpha energy levels may correlate well with performance for some test subjects and not others, and that alpha levels may correlate well with performance during some portions of a test and not others. Delta energy levels were also inconsistent with failures (missed events) because there were decreases for some subjects and increases for others. Beta energy showed relatively little sensitivity during failures in the test subject's for this study. The inconsistencies discovered in the various traditional frequency bands in this work may also help explain the possible inconsistencies reported in scoring stage W to stage 1 sleep. Of course, when scoring sleep segments, the data to be scored is in 15 or 30 second epochs and not 2 second clips. This may certainly account for some of the variations which could be averaged out in the larger data segments. Small segments are used in this work because the ultimate goal is the development of a detection device and small data segments are preferred to increase the responsiveness of a drowsiness measure observed from the data. Response times on the order of a few seconds are desirable for such a system. Looking at the high frequency bands, we can see that all of the high frequency bands shown consistent decreases during failures. The largest changes occur in the higher frequencies as shown. The high frequency band appears to be much more sensitive than the lower frequency bands with relative changes on the order of 10-times and greater from hits to misses. Our analysis has clearly demonstrated that the high frequency range has a definite correlation with extreme drowsiness and has the potential for improved sensitivity over the traditional frequency bands. For these reasons, the remaining research will focus on the high frequency activity. As a quick recap: We now have the necessary justification to fully pursue high frequency analysis.
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2010 Consolidated Research of Richmond, Inc. All rights reserved. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||